- Microscope: FEI TITAN Krios - Detector: Gatan K3 BIOQUANTUM - Acceleration voltage (kV): 300 - The pixel size in the image header (1 Å/px) is wrong. 0.83 Å/px is the calibrated value. - Spherical aberration (Cs) : 2.7 - Nominal Defocus (μm): 0.8 - 2.0 - Average electron dose per image (e-/A2): 50.8 - Please use 'gain.mrc' as a gain reference without rotation and y-flipping.
- Microscope: FEI TITAN Krios - Detector: Gatan K3 BIOQUANTUM - Acceleration voltage (kV): 300 - The pixel size in the image header (1 Å/px) is wrong. 0.83 Å/px is the calibrated value. - Spherical aberration (Cs) : 2.7 - Nominal Defocus (μm): 0.8 - 2.0 - Average electron dose per image (e-/A2): 50.2 - Please use 'gain.mrc' as a gain reference without rotation and y-flipping.
- Microscope: FEI TITAN Krios - Detector: Gatan K3 BIOQUANTUM - Acceleration voltage (kV): 300 - The pixel size in the image header (1 Å/px) is wrong. 0.83 Å/px is the calibrated value. - Spherical aberration (Cs) : 2.7 - Nominal Defocus (μm): 0.8 - 2.0 - Average electron dose per image (e-/A2): 50.0 - Please use 'gain.mrc' as a gain reference without rotation and y-flipping.
- Microscope: FEI TITAN Krios - Detector: Gatan K3 BIOQUANTUM - Acceleration voltage (kV): 300 - The pixel size in the image header (1 Å/px) is wrong. 0.83 Å/px is the calibrated value. - Spherical aberration (Cs) : 2.7 - Nominal Defocus (μm): 0.8 - 2.0 - Average electron dose per image (e-/A2): 50.0 - Please use 'gain.mrc' as a gain reference without rotation and y-flipping.