Unaligned movie frames of the individual tilt exposures. Each movie contains 6 frames and was exposed with 1.97 electrons per Å^2, or 0.3 electrons per Å^2 per frame. Movies are compressed and require the gainref file "CountRef_TS_TS_test_02_000_0.0.mrc" (uploaded in a separate directory) for motion correction. Each file has an .mdoc file associated that describes details of the exposure. These are found in the folder "mdoc_individual_tilts". In addition, the folder "tomostar" contains a .tomostar file for each tilt series that contains the information on tilt series alignment in Warp format (version 1.09). The folder "gainref" possesses the gain reference file to unpack and motion correct the tilt movie frames.
Each tilt series contains 61 tilt images, collected from -60 to +60 degrees in two degree increments, using the dose-symmetric Hagen scheme. The folder also contains SerialEM .log files with details about tilt series aquistion, and Serial .mdoc files with details about the tilt series.
Reconstructed subtomograms used for the final reconstruction, binned to 3Å/px al at bnox size of 180px. The folder also contains 3D ctf subvolumes for each subtomogram. The folder "star_file" contains a RELION (v3.1) star file that contains the refined orientations of each subtomogram.